Infrared sensor
Accurate data
High Efficiency
Intelligent control
Safe and reliable
Space saving
Powerful functions
The pre-conditioned specimen is clamped in the test cell, nitrogen with stable relative humidity flows on one side of the specimen while a stream of dry nitrogen flows on the other side. Due to the humidity difference, water vapor permeates through the specimen from the high humidity side into the low humidity side, and is carried to the infrared sensor by the dry nitrogen flow. By analyzing the electrical signals generated by the water vapor, the sensor calculates the water vapor concentration and the water vapor transmission rate.
ASTM F1249ãISO 15106-2ãGB/T 26253ãJIS K7129ãYBB00092003-2015
Applications |
---|
|
|
Table 1: Test Parameters Note2
Parameters/Model | C306H | |
---|---|---|
Test range | g/(m2·day) (Standard area 50cm2) | 0.02ï40 |
g/(m2·day) (MASK area 5cm2) | 0.2ï400ï¼Optionalï¼ | |
g/(m2·day) (MASK area 1cm2) | 1ï2000ï¼Optionalï¼ | |
Resolution | g/(m2·day) | 0.0001 |
Repeatability | g/(m2·day) | 0.02 or 2%ï¼take the greater |
Test Temperature | | 15ï50ï¼5ï60ï¼Optionalï¼ |
Temperature fluctuation | | ±0.15 |
Test Humidity | %RHï¼Within standard temperature rangeï¼ | 100%ï¼5ï902% |
Additional Functions | DataShieldTM note3 | Optional |
GMP Computer System requirement | Optional | |
CFR21Part11 | Optional |
Table 2: Technical Specifications
Test Cell | 6 Cells |
---|---|
Specimen Size | 4.6 x 4.6ï¼11.7cm11.7cmï¼ |
Specimen Thickness | 120 Milï¼3mmï¼ |
Standard test area | 50cm2 |
Carrier Gas | 99.999% High purity nitrogen(Outside of supply scope) |
Carrier Gas Pressure | 40.6 PSI / 280 kPa |
Port Size | 1/8 Metal tube |
Instrument Dimension | 23.6 H x 19.2 W x 25.9 D (60cm 49cm 66cm) |
Power Supply | 120VAC±10% 60Hz / 220VAC±10% 50Hzï¼one of twoï¼ |
Net Weight | 220Lbsï¼100kgï¼ |
Table 3: Product Configuration
Standard Configuration | Instrument mainframe, tablet, sampler, vacuum grease, Î6 mm PU tubing |
---|---|
Optional Parts | Air compressor, CFR21Part11, GMP Computer System Requirement, DataShieldTM note3 |
Note | The gas supply port of the instrument is Î6 mm PU tubingï¼pressure79.7 PSI / 550 kPaï¼, customers need to prepare gas supply. |
Note 1: The described product functions are subject to the specification in "Technical Parameters".
Note 2: The parameters in the table are measured in Labthink laboratory by professional operators according to the requirements and conditions stapulated in laboratory environmental standards.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.