Innovative structure design & upgraded automatic operation
Upgraded testing accuracy and efficiency
Excellent temperature and pressure control technique
Versatile functions based on standards, customization available
High end, secure and easy-to-use embedded computer controlled system
C130H is designed in accordance with differential pressure method. The pre-conditioned specimen is mounted in the gas diffusion cell as to form a sealed barrier between two chambers. The lower-pressure chamber is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber and a constant pressure difference is generated between the two chambers. The gas permeates through the specimen from higher pressure side into the lower pressure side. The gas permeability and other barrier properties of the specimen can be obtained by monitoring the pressure changes in the lower chamber.
ISO 2556, ISO 15105-1, GB/T 1038-2000, ASTM D1434, JIS K7126-1, YBB 00082003
This instrument is applicable to the determination of gas permeability of:
Basic Applications |
---|
|
|
Extended Applications |
|
|
|
|
|
|
|
|
|
Specifications | Film Test |
---|---|
Test Range | 0.01 50,000 cm3/m2·24h·0.1MPa |
Resolution | 0.001 cm3/ m2·24h·0.1MPa |
Temperature Range | 10C 50C (room temperature 23C) |
Resolution | 0.01C |
Temperature Fluctuation | ±0.05C |
Temperature Accuracy | ±0.3C (calibration port) |
Vacuum Resolution | 0.01Pa |
Vacuum Accuracy | Displayed Value ±0.2% (1%100% of sensor range) |
Vacuum Degree of Test Chamber | < 10 Pa |
Number of Specimens | 3 with independent test results Customization available for other number of specimens |
Specimen Size | Î97 mm |
Test Area | 38.48 cm2 |
Test Gas | O2, N2, and CO2 (outside of supply scope) Humidifying of Test Gas (customization available) |
Test Pressure | 10KPa 210KPa |
Gas Supply Pressure | 0.5 Mpa 0.6 Mpa (73psi 87psi ) |
Port Size | Î6 mm PU Tubing |
Instrument Dimension | 710 mm (L) 350 mm (W) 630 mm (H) |
Power Supply | 220VAC±10% 50Hz / 120VAC±10% 60Hz |
Net Weight | 100 kg |
Note 1: The described test standard, applications and product features should be in line with Technical Specifications.
Note 2: The parameters in the table are measured by professional operators in Labthink laboratory under strictly controlled laboratory conditions.
Note 3: DataShieldTM provides safe and reliable data application support. Multiple Labthink instruments can share one single DataShieldTM system which can be configured as required.
LABTHINK INSTRUMENTS CO. LTD.
All Rights Reserved.(Terms of Use) Developed and Managed by Infocom Network Private Limited. |