Method A:the air flow is stabilized vertically through the sample at a constant flow. Then, the pressure difference formed on both sides of the sample is measured under this condition, and air flow resistance and other parameters are calculated.
Method B:through adjustment, a constant pressure difference is formed on both sides of the sample, the air flow passing vertically through the given area of the sample in a certain time is measured, and the air permeability and other parameters are calculated.
ISO 9237, ISO 4638, ISO 5636, GB/T 10655, GB/T 5453, GB/T 4689.22, GB/T 13764, ASTM D737, TAPPI T460, JIS P8117
Basic Applications |
---|
|
|
|
|
Extended Applications |
|
|
Table 1: Test Parameters Note2
Parameters/Model | TQD-G1A | |
---|---|---|
Differential Pressure Test Range | Pa | 0ï1000 |
Differential Pressure Resolution | Pa | 0.1 |
Differential Pressure Accuracy | Pa | 0.25% |
Flow Measurement Range | L/min | 0ï30 |
Flow Resolution | L/min | 0.1 |
Flow Accuracy | L/min | 2% |
Table 2: Technical Specifications
No. of Test Cell | 1 Cell |
---|---|
Specimen Size | 2 x 2 (5cm5cm) |
Specimen Thickness | 40 Mil (1mm) |
Test Area | 20cm2(other sizes can be customized) |
Gas Specification | Clean Compressed Air (the air source is provided by the user) |
Air Source Pressure | 72.5 PSI / 500kPa |
Interface Size | Î8 mm Polyurethane Pipe |
Instrument Dimension | 12 H x 18 W x 20 D (30cm 46cm 52cm) |
Power Supply | 120VAC±10% 60Hz / 220VAC±10% 50Hz (one from the two) |
Net Weight | 55Lbs (25kg) |
Table 3: Product Configuration
Standard Configuration | Mainframe, one set of fixtures and Î8mm polyurethane pipe |
---|---|
Optional Parts | Computer software, air compressor |
Notes | The compressed air inlet of this tester isÎ8mm polyurethane pipe (pressure79.7 PSI / 550 kPa, outlet flow > 120L/min); Air source is to be provided by the user. |
Note 1: the reference standards, test applications and product features mentioned are subject to the parameters in the "Technical Parameters".
Note 2: the parameters in the table are measured by professional operators in Labthink laboratory according to the requirements and conditions of relevant laboratory environmental standards.
LABTHINK INSTRUMENTS CO. LTD.
All Rights Reserved.(Terms of Use) Developed and Managed by Infocom Network Private Limited. |